Proof-of-concept system for label-free cell measurements

Developing a low-cost, label-free cell characterisation system

As the industry looks to establish faster automation of advanced therapy manufacturing, there is a growing need for accurate, real-time insights throughout the production process. These in-process analytics are essential to help therapy manufacturers make timely, informed decisions for improved process adaptive control or product quality, saving significant time and costs during manufacture.

Team Consulting’s applied science experts set out to explore how miniaturised and label-free analytical technologies could be rapidly integrated into a system to provide rich, in-process data at a low cost. Combining microfluidics, impedance spectroscopy and optical sensing, we developed a proof-of-concept system capable of characterising individual cells and generating meaningful process data closer to manufacture, enabling deeper process understanding and supporting future adaptive control.

Stainless Steel Biopharmaceutical Bioreactor Vessel in Cleanroom Facility

What are the challenges of cell therapy manufacturing?

Ex vivo cell therapy manufacturing remains difficult to scale because it involves complex, variable biological starting materials and multi-step processes that often rely on manual handling and offline testing. Although automation can improve consistency, many existing analytical methods are not designed for the small batch sizes, time-sensitive workflows and aseptic constraints of advanced therapies. As a result, manufacturers often lack the real-time process data needed to link critical process parameters to product quality, identify poor runs early and move towards more adaptive control.

Gloved hands working on microfluidics prototype analytics device

Rapidly designing and integrating in-process analytics

Speed of manufacture and cost are both major factors in cell and gene therapy manufacturing. We set out to demonstrate that simple, low cost, compact analytical devices can be rapidly designed, assembled and integrated into cell and gene therapy manufacturing workflows to provide powerful in process insights.

To minimise development time and cost, we followed a pragmatic “buy, adapt and make” strategy. This involved purchasing reliable, complex components off the shelf where suitable, and adapting existing equipment where modifications could meet the project requirements. Custom components were developed only where an appropriate solution was unavailable. This approach allowed us to focus engineering time and effort on the fluidic, optical, electronic and software interfaces needed to integrate the system and capture live impedance flow cytometry data and video data.

The resulting system enables particles of different characteristics to flow through a tightly controlled sensing volume where optical and electrical measurements can be captured simultaneously and provide real-time cell characteristics.

Microfluidics analytics prototype device
Microfluidics analytics prototype device

De-risking development for rapid implementation

A key focus throughout the development was to reduce development risk and determine technical feasibility early on. Using rapid prototyping techniques, including custom-designed interfaces and 3D-printed components, in just three weeks our applied science experts assembled and validated a functioning analytical system. By demonstrating that the approach could reliably distinguish between simple, predictable plastic particles and generate meaningful data, we established confidence in the underlying technique before progressing on to biological cells.

This rapid, evidence-driven approach enabled us to move quickly from concept generation to functional testing, while maintaining flexibility to refine the system architecture as new insights emerged.

Developing a proof-of-concept system for label-free cell measurements

The proof-of-concept system was designed for label-free measurements, able to count and differentiate cells such as white blood cells in real time without needing to pre-condition them. To test the setup and prove the principle, we selected two sets of spherical latex particles of different sizes comparable to human white blood cells (WBCs). Two sensing methods were implemented to test their effectiveness: optical brightfield imaging and impedance flow cytometry.

Applying optical brightfield imaging and impedance flow cytometry sensing

Developing a miniaturised optical brightfield imaging system

Optical brightfield imaging is a microscopy technique that captures images of cells under visible light, allowing them to be identified, classified and counted based on their size, shape, morphology and density. Our device engineers built this functionality into the system using predominantly off-the-shelf components, providing a cost-effective and scalable alternative to custom optics while delivering sufficient performance for the application.

To enable efficient optimisation of the optical design and component selection, we combined ray-tracing simulations in Ansys Zemax with rapid bench testing in our optics laboratory. As the system was designed to be integrated at-line within different automated manufacturing platforms, we adapted the technology into a compact 10 × 10 cm footprint using custom 3D-printed fixtures.

Beyond generating complementary measurement data, the optical system also played a key role in the development by providing visual confirmation of particle behaviour within the microfluidic channel, helping validate electrical signals, optimise operating parameters such as flow rate, and identify opportunities for future design improvements.

Impedance flow cytometry

Impedance flow cytometry counts and characterises cells in a laminar flow regime, by measuring how each cell perturbs an applied electric field as it passes between microelectrodes. In simple terms, the cell, electrodes and conductive fluid can be modelled as an electrical circuit, so changing the measurement frequency probes different electric properties of the system, including the cell. Here we operate at frequencies where the current cannot penetrate the cell membrane, meaning the measured signal primarily reflects cell size and morphology. At higher frequencies, however, the current can pass through the membrane and interact with the cell’s internal structures, providing additional information about its composition and properties.

We implemented this within the system to provide rapid, label-free characterisation of individual particles within a microfluidic environment. Using an Analog Devices (ADI) evaluation board and impedance module integrated with a custom microfluidic interface, the system captures electrical signatures as particles passed through the sensing region. We then developed bespoke software to visualise and log impedance data in real time, enabling rapid assessment of system performance and measurement quality.

Calibrating the system for impedance measurement

Using a syringe pump, we were able to generate stable, programmable flow profiles, while an off-the-shelf microfluidic chip, with channel dimensions suited to the test particles, enabled controlled flow through the sensing region. To maximise measurement sensitivity, the system was carefully calibrated and operated within the performance limits of the impedance measurement electronics. Signal quality was improved through careful cable management, minimised electrical interference and optimisation of acquisition parameters. To further reduce measurement artefacts, fluids were degassed prior to use and the fluidic architecture was designed to minimise bubble formation. Flow conditions were selected to ensure steady laminar flow and make sure that particles remained within the sensing region for sufficient time to be accurately detected and characterised.

The result was an adaptable system capable of producing impedance-based measurements within a microfluidic environment and generating data capable of distinguishing between latex particles of different sizes.

Proof-of-concept system for label-free cell measurements

Outcome

We developed a functioning proof-of-concept system that demonstrates the potential to integrate optical and impedance-based measurements within a compact microfluidic platform. The system showcases how label-free analytical technologies could support future applications in cell health monitoring, process understanding and adaptive manufacturing control, with the potential for future development and adaptations to different use cases.

By generating measurements closer to the process itself, systems such as these could provide manufacturers with earlier insight into cell quality and process performance, support more informed and timely decision-making and reduce reliance on end-point testing.

How can we help you?

Let’s start working together on your next challenge.